Hybrid Closed-Loop Insulin Delivery in Type 1 Diabetes During Supervised Outpatient Conditions
Benyamin Grosman1, Jacob Ilany2, Anirban Roy3
1Medtronic MiniMed, Northridge, CA, USA benyamin.grosman@medtronic.com.
Background:
Efficacy and safety of the Medtronic Hybrid Closed-Loop (HCL) system were tested in subjects with type 1 diabetes in a supervised outpatient setting.
Methods:
The HCL system is a prototype research platform that includes a sensor-augmented insulin pump in communication with a control algorithm housed on an Android-based cellular device. Nine subjects with type 1 diabetes (5 female, mean age 53.3 years, mean A1C 7.2%) underwent 9 studies totaling 571 hours of closed-loop control using either default or personalized parameters. The system required meal announcements with estimates of carbohydrate (CHO) intake that were based on metabolic kitchen quantification (MK), dietician estimates (D), or subject estimates (Control). Postprandial glycemia was compared for MK, D, and Control meals.
Results:
The overall sensor glucose mean was 145 ± 43, the overall percentage time in the range 70-180 mg/dL was 80%, the overall percentage time <70 mg/dL was 0.79%. Compared to intervals of default parameter use (225 hours), intervals of personalized parameter use (346 hours), sensor glucose mean was 158 ± 49 and 137 ± 37 mg/dL (P < .001), respectively, and included more time in range (87% vs 68%) and less time below range (0.54% vs 1.18%). Most subjects underestimated the CHO content of meals, but postprandial glycemia was not significantly different between MK and matched Control meals (P = .16) or between D and matched Control meals (P = .76). There were no episodes of severe hypoglycemia.
Conclusions:
The HCL system was efficacious and safe during this study. Personally adapted HCL parameters were associated with more time in range and less time below range than default parameters. Accurate estimates of meal CHO did not contribute to improved postprandial glycemia.
More Related Videos
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Diabetes Mellitus: Type 2 and Gestational
Insulin: Biosynthesis, Chemistry, and Preparation
Damage or functional impairment of β-cells inhibits insulin production, leading to diabetes. Diabetes treatment...