Dose response characteristics of a novel CCD camera-based electronic portal imaging device comparison with OCTAVIUS
Akbar Anvari1, Seyed Mahmoud Reza Aghamiri, Seyed Rabie Mahdavi
1Department of Radiation Medicine Engineering, Shahid Beheshti University, Tehran, Iran.
Aim:
Dosimetric properties of a CCD camera-based Electronic Portal Imaging Device (EPID) for clinical dosimetric application have been evaluated. Characteristics obtained by EPID also compared with commercial 2D array of ion chambers.
Materials And Methods:
Portal images acquired in dosimetry mode then exported raw fluence or uncorrected images were investigated. Integration time of image acquisition mode has adjusted on 1 s per frame.
Results:
As saturation of camera of the EPID, dose response does not have linear behavior. The slight nonlinearity of the camera response can be corrected by a logarithmic expression. A fourth order polynomial regression model with coefficient of determination of 0.998 predicts a response to absolute dose values at less than 50 cGy. A field size dependent response of up to 7% (0.99-1.06) relative OCTAVIUS detector measurement was found. The EPID response can be fitted by a cubic regression for field size changes, yielded coefficient of determination of 0.999.
Conclusions:
These results indicate that the EPID is well suited for accurate dosimetric purposes, the major limitation currently being due to integration time and dead-time in frame acquisition.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...