The Use of Sound Level Meter Apps in the Clinical Setting
Purpose:
The purpose of this study was to compare sound level meter (SLM) readings obtained using a Larson-Davis (Depew, NY) Model 831 Type 1 SLM, a RadioShack (Fort Worth, TX) SLM, and iPhone 5 (Apple, Cupertino, CA) SLM apps.
Method:
In Procedure 1, pure tones were measured in an anechoic chamber (125, 250, 500, 1000, 2000, 4000, and 8000 Hz); sound pressure levels (SPLs) ranged from 60 to 100 dB SPL in 10-dB increments. In Procedure 2, human voices were measured. Participants were 20 vocally healthy adults (7 women, 13 men; mean age = 25.1 years). The task was to sustain a vowel "ah" at 3 intensity levels: soft, habitual, and loud. Microphones were lined up equal distances from the participant's mouth, and recordings were captured simultaneously.
Results:
Overall, the 3 SLM apps and the RadioShack SLM yielded inconsistent readings compared with the Type 1 SLM.
Conclusion:
The use of apps for SPL readings in the clinical setting is premature because all 3 apps adopted were incomparable with the Type 1 SLM.
More Related Videos
Related Concept Videos
Sound Intensity Level
The human ear can perceive an extensive range of sound intensity, necessitating the use of the logarithmic scale to define a physical quantity—the intensity level. It is a ratio of two intensities and...
Applications of Logarithms
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...