Related Experiment Video
Updated: Mar 25, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Spot-scanned pancreatic stereotactic body proton therapy: A dosimetric feasibility and robustness study
Terence T Sio1, Kenneth W Merrell2, Chris J Beltran2
1Department of Radiation Oncology, Mayo Clinic, 200 First St. SW, Rochester, MN 55901, USA; Department of Radiation Oncology, Mayo Clinic Hospital, 5777 E. Mayo Blvd., Phoenix, AZ 85054, USA.
Purpose:
We explored the dosimetric potential of spot-scanned stereotactic body proton therapy (SBPT) for pancreatic cancer.
Methods:
We compared SBPT to stereotactic body intensity-modulated radiotherapy (SB-IMRT) in 10 patients. We evaluated 3 variables in SBPT planning: (1) 4 and 6 mm spot size; (2) single vs. multi-field optimization (SFO vs. MFO); and (3) optimization target volume (OTV) expansion. Robustness analysis was performed with unidirectional isocenter shifts of ±3 mm in x, y, and z and ±3% stopping power uncertainties.
Results:
SBPT plans had lower V10Gy for the stomach and small and large bowels. Under static robustness, a 5 mm OTV and SFO-6 mm spot size represented the best compromise between target and normal structure. A 4-mm spot-size and 3 mm OTV resulted in significant target underdosing with deformable dose accumulation analysis.
Conclusions:
This study provides a critical basis for clinical translation of spot size, optimization technique, and OTV expansion for pancreatic SBPT.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...