Related Experiment Video
Updated: Mar 25, 2026

Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Radiation intensity (CTDIvol) and visibility of anatomical structures in head CT examinations
Sameer Tipnis1, Rajesh Thampy, Zoran Rumboldt
1Medical University of South Carolina. tipnis@musc.edu.
Abstract:
The purpose of this study was to quantify how changing the amount of radiation used to perform routine head CT examinations (CTDIvol) affects visibility of key anatomical structures. Eight routine noncontrast head CT exams were selected from six CT scanners, each of which had a different CTDIvol setting (60 to 75 mGy). All exams were normal and two slices were selected for evaluation, one at the level of basal ganglia and the other at the fourth ventricle. Three experienced neuroradiologists evaluated the visibility of selected structures, including the putamen, caudate nucleus, thalamus, internal capsule, grey/white differentiation, and brainstem. Images were scored on a five-point scoring scheme (1, unacceptable, 3, satisfactory, and 5, excellent). Reader scores, averaged over the cases obtained from each scanner, were plotted as a function of the corresponding CTDIvol. Average scores for the fourth ventricle were 3.06 ± 0.83 and for the basal ganglia were 3.20 ± 0.86. No image received a score of 1. Two readers showed no clear trend of an increasing score with increasing CTDIvol. One reader showed a slight trend of increasing score with increasing CTDIvol, but the increase in score from a 25% increase in CTDIvol was a fraction of the standard deviation associated average scores. Collectively, results indicated that there were no clear improvements in visualizing neuroanatomy when CTDIvol increased from 60 to 75 mGy in routine head CT examinations. Our study showed no apparent benefit of using more than 60 mGy when performing routine noncontrast head CT examinations.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography