The Effect of Changing Scan Mode on Trabecular Bone Score Using Lunar Prodigy
Weiwen Chen1, Anthony Slattery2, Jacqueline Center1
1Garvan Institute of Medical Research, St Vincent's Hospital, Sydney, Australia.
Summary
Trabecular bone score (TBS) measurements on GE Lunar Prodigy machines differ based on scanning mode. Thick mode acquisitions yield significantly higher TBS values than standard mode, impacting fracture risk assessment.
Area of Science:
- Osteoporosis Research
- Medical Imaging Analysis
- Bone Densitometry
Background:
- Trabecular bone score (TBS) assesses bone microarchitecture and predicts fracture risk.
- TBS is increasingly used in osteoporosis assessment and integrated into FRAX®.
- GE Lunar DXA machines utilize different acquisition modes (thin, standard, thick) based on patient anatomy.
Purpose of the Study:
- To investigate the impact of different acquisition modes on TBS values.
- To compare TBS measurements from standard and thick modes on the GE Lunar Prodigy.
- To determine if scanning mode affects TBS in osteoporosis assessment.
Main Methods:
- Retrospective review of 30 patients with paired lumbar spine DXA scans (standard and thick modes).
- Scans performed on GE Lunar Prodigy on the same day without repositioning.
- Analysis of lumbar spine bone mineral density (L1-L4) and TBS from paired scans.
Main Results:
- No significant difference in lumbar spine bone mineral density between standard and thick modes.
- Significantly higher TBS values were observed in thick mode compared to standard mode (mean difference: 0.24, 20%).
- TBS values demonstrated a statistically significant difference based on the acquisition mode used.
Conclusions:
- Preliminary data indicate that TBS values from GE Lunar Prodigy are dependent on the scanning acquisition mode.
- Further research is needed to understand the cause of this discrepancy.
- Development of appropriate protocols is required to ensure consistent TBS measurements across different scanning modes.