Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
Published on: April 4, 2013
Magnetic resonance imaging-compatible circular mapping catheter: an in vivo feasibility and safety study
Delphine Elbes1,2, Julie Magat1, Assaf Govari3
1IHU LIRYC/CRCTB, INSERM U1045, University of Bordeaux, Bordeaux, France.
Aims:
Interventional cardiac catheter mapping is routinely guided by X-ray fluoroscopy, although radiation exposure remains a significant concern. Feasibility of catheter ablation for common flutter has recently been demonstrated under magnetic resonance imaging (MRI) guidance. The benefit of catheter ablation under MRI could be significant for complex arrhythmias such as atrial fibrillation (AF), but MRI-compatible multi-electrode catheters such as Lasso have not yet been developed. This study aimed at demonstrating the feasibility and safety of using a multi-electrode catheter [magnetic resonance (MR)-compatible Lasso] during MRI for cardiac mapping. We also aimed at measuring the level of interference between MR and electrophysiological (EP) systems.
Methods And Results:
Experiments were performed in vivo in sheep (N = 5) using a multi-electrode, circular, steerable, MR-compatible diagnostic catheter. The most common MRI sequences (1.5T) relevant for cardiac examination were run with the catheter positioned in the right atrium. High-quality electrograms were recorded while imaging with a maximal signal-to-noise ratio (peak-to-peak signal amplitude/peak-to-peak noise amplitude) ranging from 5.8 to 165. Importantly, MRI image quality was unchanged. Artefacts induced by MRI sequences during mapping were demonstrated to be compatible with clinical use. Phantom data demonstrated that this 10-pole circular catheter can be used safely with a maximum of 4°C increase in temperature.
Conclusions:
This new MR-compatible 10-pole catheter appears to be safe and effective. Combining MR and multipolar EP in a single session offers the possibility to correlate substrate information (scar, fibrosis) and EP mapping as well as online monitoring of lesion formation and electrical endpoint.
More Related Videos
09:13Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging