Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
Published on: January 16, 2019
Comparison of different methods for measurement of electrolytes in patients admitted to the intensive care unit
Sinan Yilmaz1, Hilal B Uysal, Mücahit Avcil
1Department of Anesthesiology and Reanimation, Adnan Menderes University School of Medicine, Aydin, Turkey. E-mail. dr_snnylmz@hotmail.com.
Objectives:
To investigate whether electrolyte levels measured by using blood gas analyzers (ABG) and auto-analyzers (AA) are equivalent and can be used interchangeably.
Methods:
This observational prospective study was conducted in 100 patients admitted to the Intensive Care Unit, Adnan Menderes University School of Medicine, Aydin, Turkey, between March and August 2014. Samples for both AA and ABG analyzers were collected simultaneously from invasive arterial catheters of patients. The electrolyte levels were measured by using 2 methods.
Results:
The mean sodium level measured by ABG was 136.1 ± 6.3 mmol/L and 137.8 ± 5.4 mmol/L for AA (p=0.001). The Pearson's correlation coefficient was 0.561 (p less than 0.001). The Bland-Altman 95% limits of agreement were -9.4 to 12.6 mmol/L. The mean potassium levels measured by ABG was 3.4 ± 0.7 mmol/L and AA was 3.8 ± 0.7 mmol/L (p=0.001). The Bland-Altman comparison limits were -0.58 to 1.24 and the associated Pearson's correlation coefficient was 0.812 (p less than 0.001).
Conclusion:
The results of the 2 analyzing methods, in terms of sodium, were not equivalent and could not be used interchangeably. However, according to the statistical analyses results, by including, but not blindly trusting these findings, urgent and vital decisions could be made by the potassium levels obtained from the BGA, but a simultaneous follow-up sample had to be sent to the central laboratory for confirmation.
Related Concept Videos
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Interfacial Electrochemical Methods: Overview
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Controlled-Current Coulometry: Overview
Coulometry: Overview