A Human Corneal Organ Culture Model of Descemet's Stripping Only with Accelerated Healing Stimulated by Engineered Fibroblast Growth Factor 1
Published on: July 22, 2022
5.2K
Polarization of human donor corneas
Mohit Parekh1, Alessandro Ruzza2, Stefano Ferrari2
1International Center for Ocular Physiopathology (ICOP), The Veneto Eye Bank Foundation, Via Paccagnella 11, 30174, Mestre, Venice, Italy. mohit.parekh@fbov.it.
Cell and Tissue Banking
|February 28, 2016
Summary
This study used a polarizing device to assess human donor corneas, finding significant changes in polarization power and cross patterns with rotation. These findings suggest polarization analysis could evaluate corneal tissue quality for DSAEK grafts.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Corneal Science
Background:
- Descemet's stripping automated endothelial keratoplasty (DSAEK) is a common corneal transplantation technique.
- Assessing graft orientation and tissue quality is crucial for successful outcomes.
- Objective methods to evaluate corneal optical properties are needed.
Purpose of the Study:
- To investigate the de-orientation effect on human donor corneas using a polarizing device (Lumaxis®).
- To evaluate the polarization power and cross patterns of corneas during simulated DSAEK graft preparation.
- To explore the potential of polarization analysis as an objective measure of corneal tissue quality.
Main Methods:
- Forty human donor corneas were analyzed using a Lumaxis® polarizing device.
- Corneas were subjected to simulated graft preparation, including cuts and rotations up to 45°.
- Polarization power and cross patterns were monitored and recorded at various stages.
Main Results:
- A statistically significant difference (p < 0.05) in polarizing power was observed with angular changes.
- Corneal cross patterns deformed and lost sharpness as the tissue was rotated by 45°.
- The Lumaxis® device identified poor quality tissue based on polarization characteristics.
Conclusions:
- Polarization analysis with Lumaxis® can detect changes in corneal tissue during simulated DSAEK preparation.
- Observed changes in polarization power and cross patterns may serve as objective parameters for evaluating corneal optical properties.
- Further research is needed to establish the clinical relevance of these polarization changes for visual acuity in patients receiving DSAEK grafts.