Simultaneous Distinction of Monospecific and Mixed DFS70 Patterns During ANA Screening with a Novel HEp-2 ELITE/DFS70 Knockout Substrate
Published on: January 17, 2018
Detection of anti-dsDNA antibodies by computer-aided automated immunofluorescence analysis
Gabriella Lakos1, Marlene Gonzalez1, David Flaherty1
1Inova Diagnostics, Inc, San Diego, CA, USA.
Introduction:
NOVA View is a computer aided fluorescence microscope that is used for the automated reading and interpretation of indirect immunofluorescent tests in diagnostic immunology. The objective of the present study was to evaluate the performance of the NOVA View® system for the measurement of anti-dsDNA antibodies using the Crithidia luciliae indirect immunofluorescence test (CLIFT) technology.
Methods:
Analytical performance of NOVA View CLIFT was assessed in repeatability (within run) and reproducibility (between runs and instruments) studies. Two hundred-fifty patient samples (N=200 consecutive samples and N=50 samples from systemic lupus erythematosus patients) were tested to evaluate the agreement between results generated with NOVA View CLIFT, and those obtained with manual microscopic reading of the same slides. Positivity rate in SLE was assessed on the 50 SLE samples.
Results:
The NOVA View system showed high level of repeatability and reproducibility within runs, between runs, and between instruments. Agreement of NOVA View software interpretation and digital image reading results with manual microscopic reading results was 96.0%, and the same positivity rate was obtained on SLE samples by NOVA View digital image reading as that of manual microscopic reading (36.0% vs. 38.0%, respectively).
Conclusion:
Results generated by NOVA View CLIFT were equivalent to those obtained by manual microscopic reading on a large routine sample set. NOVA View demonstrated consistency within and between runs, and between instruments. Automation of CLIFT provides reliability and is a suitable alternative for routine clinical laboratories.