Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
[Accuracy of three common optometry methods in examination of refraction in juveniles]
Ting Su1, Xiaoshan Min1, Shuangzhen Liu1
1Department of Ophthalmology, Xiangya Hospital, Central South University, Changsha 410008, China.
Objective:
To compare the results of the three methods of Suresight handheld autorefractor, table-mounted autorefractor and retinoscopy in examination of juveniles patients with or without cycloplegia.
Methods:
Firstly, 156 eyes of 78 juveniles (5 to 17 years old) were examined by using WelchAllyn Suresight handheld autorefractor and NIDEK ARK-510A table-mounted autorefractor with or without cycloplegia; secondly, retinoscopy was performed with cycloplegia.
Results:
The spherical power measured by methods without cycloplegia were significantly greater than those measured with cycloplegia (P<0.05); without cycloplegia, there was no significant difference in spherical power, cylindrical power and cylindrical axis between Suresight handheld autorefractor and retinoscopy (P>0.05). These results were highly consistent, suggesting a tendency towards a short sight. However, the spherical power and cylindrical power measured by table-mounted autorefractor was significantly different (P<0.05); with cycloplegia, there was significant difference in spherical power between Suresight handheld autorefractor and retinoscopy (P<0.05).
Conclusion:
Cycloplegic retinoscopy is necessary for juvenile refraction examination. Under natural pupil situation, Suresight handheld autorefractor is better than table-mounted autorefractor, though both show a myopia tendency. Nevertheless, table-mounted autorefractor can be taken as a recommendation for the prescription of lens trial. As a strong reference for subjective optometry, retinoscopy should be the gold standard for measuring refractive errors.