Related Experiment Videos
Insulin delivery by implanted pumps: one-year trial with programmable infusion systems (the Point Study).
1IIIrd Medical Department, City Hospital Munich, FRG.
The International Journal of Artificial Organs
|December 1, 1989
Summary
This study assessed stabilized insulin in implanted pumps for type I diabetes management. Results show safe and effective glucose control, despite some technical challenges with the remote-controlled devices.
Area of Science:
- Biomedical Engineering
- Endocrinology
- Diabetes Technology
Background:
- Type I diabetes requires continuous insulin management.
- Implantable insulin pumps offer potential for improved glycemic control.
- Stabilized insulin formulations are crucial for device compatibility.
Purpose of the Study:
- To evaluate the feasibility, safety, and efficacy of a stabilized insulin preparation (HOE 21GH) in remote-controlled implanted infusion pumps.
- To assess glycemic control and complication rates in type I diabetic patients using this system.
Main Methods:
- A feasibility trial involving twenty type I diabetic patients with implanted intraperitoneal or intravenous insulin pumps.
- Collection of 36,077 self-monitored blood glucose measurements over 18.2 patient years.
- Monitoring of symptomatic hypoglycemia, HbA1c levels, and device-related complications.
Main Results:
- 62.9% of blood glucose measurements fell within the target range (70-160 mg/dl).
- Symptomatic hypoglycemia occurred infrequently (2.6 per patient-month), with 0.22 per patient-year requiring medical attention.
- Median HbA1c levels were within or slightly above the upper normal range; three pumps required explantation due to surgical/technical issues.
Conclusions:
- Implanted, remote-controlled insulin delivery devices with stabilized insulin are clinically feasible.
- The system demonstrates potential for effective glycemic management in type I diabetes.
- Ongoing technical and clinical challenges require further investigation and optimization.