Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
Procedural characteristics of pulmonary vein isolation using the novel third-generation cryoballoon
Alexander Fürnkranz1, Fabrizio Bologna2, Stefano Bordignon2
1Cardioangiologisches Centrum Bethanien, Medizinische Klinik III, Markus Krankenhaus, Wilhelm-Epstein-Street 4, D-60431 Frankfurt am Main, Germany a.fuernkranz@gmail.com.
Aim:
A novel third-generation cryoballoon (CB3) to perform pulmonary vein isolation (PVI) has recently been released, featuring a shortened distal balloon tip when compared with the second-generation (CB2), possibly allowing for enhanced intra-ablation pulmonary vein (PV) signal mapping. We aimed to investigate procedural efficacy and safety of the CB3 as compared to the CB2.
Methods And Results:
We studied 472 consecutive patients who underwent CB-PVI for paroxysmal or persistent atrial fibrillation (CB3: 49 patients; CB2: 423 patients). Detailed procedural data and in-hospital complications were registered in a prospective database. Complete PVI using the CB only was achieved in 98% of patients in each group. Single-freeze PVI was observed in 84/88% (CB2/CB3, P = n.s.) of the PVs. Time-to-PVI (TPVI) was 49 ± 32 (CB2) and 45 ± 27 s (CB3) (P = n.s.). Time-to-PVI determination rate was higher in the CB3 group (89.5 vs. 82.6%, P = 0.016). Signal noise due to ice formation on mapping electrodes occurred after 70 ± 46 s using CB3 and did not interfere with TPVI determination. Exchange of the spiral mapping catheter with a guide wire was more frequently required in the CB3 group (8.2 vs. 0.7% patients, P < 0.001). Balloon dislodgement during hockey stick manoeuvres occurred in 6.1% patients of the CB3 group only (P = 0.001). Complication rates were not different between the groups.
Conclusion:
The CB3 offers a higher TPVI determination rate, facilitating dosing schemes based on TPVI, with equally high single-freeze efficacy compared with the CB2. The shortened distal tip of the CB3 requires adaptation of standard catheter manoeuvers to avoid balloon dislodgement.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
06:48Three-Dimensional Echocardiographic Method for the Visualization and Assessment of Specific Parameters of the Pulmonary Veins
Published on: October 28, 2020
Related Concept Videos
Venous Thrombosis III: Interprofessional Care
Varicose Veins II: Diagnostic Studies and Interprofessional Care