An inappropriate defibrillator shock during ventricular tachycardia
Mariana Faustino1, Francisco Madeira1, Carlos Morais1
1Serviço de Cardiologia, Hospital Fernando Fonseca, Amadora, Portugal.
Indian Pacing and Electrophysiology Journal
|March 4, 2016
Summary
Ventricular oversensing in patients with defibrillators can lead to inappropriate shocks. This case highlights how double-counting within the QRS complex caused a defibrillator to misinterpret slow ventricular tachycardia as ventricular fibrillation.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Device Technology
Background:
- Ventricular oversensing is a rare but serious complication in patients with implantable defibrillators.
- It can lead to inappropriate and potentially harmful therapies.
- Cardiac resynchronization-defibrillation (CRD) devices are used for heart failure and arrhythmia management.
Purpose of the Study:
- To report a case of ventricular oversensing due to double-counting in a patient with a CRD device.
- To illustrate how this oversensing led to inappropriate shock delivery.
- To emphasize the importance of device interrogation for accurate diagnosis and therapy.
Main Methods:
- Case report of a patient presenting with syncope and hyperkalemia.
- Device interrogation of a cardiac resynchronization-defibrillation system.
- Analysis of electrogram data to identify the cause of oversensing and inappropriate therapy.
Main Results:
- The patient experienced syncope attributed to hyperkalemia.
- Device interrogation revealed ventricular oversensing due to double-counting of the QRS complex during slow ventricular tachycardia.
- This led to the device misclassifying the rhythm as ventricular fibrillation and delivering an inappropriate therapy.
Conclusions:
- Ventricular oversensing, even with advanced CRD devices, remains a clinical challenge.
- Double-counting within the QRS complex is a potential mechanism for inappropriate detection.
- Thorough electrogram interrogation is crucial for diagnosing oversensing and preventing unnecessary defibrillator therapies.