Comparing hepatic 2D and 3D magnetic resonance elastography methods in a clinical setting - Initial experiences
Mikael F Forsgren1, Bengt Norén2, Johan Kihlberg2
1Wolfram MathCore AB, Linköping, Sweden; Center for Medical Image Science and Visualization (CMIV), Linköping University, SE-581 85 Linköping, Sweden; Radiation Physics, Division of Radiological Sciences, Department of Medical and Health Sciences (IMH), Linköping University, SE-581 83 Linköping, Sweden.
This pilot study demonstrates that different magnetic resonance elastography (MRE) methods yield comparable liver viscoelasticity measurements. This comparability is crucial for equipment-independent monitoring of liver fibrosis progression in clinical and research settings.
Area of Science:
- Medical Imaging
- Hepatology
- Biophysics
Background:
- Continuous monitoring of liver fibrosis is limited by the invasiveness of needle biopsies.
- Magnetic Resonance Elastography (MRE) offers a non-invasive approach for assessing liver fibrosis.
- Variations in MRE techniques necessitate validation of measurement comparability.
Purpose of the Study:
- To evaluate the comparability of viscoelastic property measurements between 2D and 3D MRE methods.
- To determine if different MRE approaches provide consistent data for liver fibrosis assessment.
Main Methods:
- Seven patients with liver disease underwent both 2D and 3D MRE on the same day.
- 2D MRE used an acoustic transducer on a GE MR system; 3D MRE used an electromagnetic transducer on a Philips MR system.
- Viscoelastic values were extracted from identical anatomical regions by an experienced radiologist.
Main Results:
- A strong correlation was observed between 2D MRE stiffness and 3D MRE elasticity (R² = 0.96, P < 0.001).
- Similarly, 2D MRE stiffness correlated well with 3D MRE 'G abs,Elastic' (R² = 0.96, P < 0.001).
- Both comparisons showed slopes close to 1 and low intercepts, indicating good agreement.
Conclusions:
- Different MRE methods can produce comparable measurements of liver viscoelastic properties.
- This comparability supports equipment-independent monitoring of liver disease progression.
- Findings are significant for both clinical practice and research in liver fibrosis.