Related Experiment Video
Updated: Mar 24, 2026

High Density Event-related Potential Data Acquisition in Cognitive Neuroscience
Published on: April 16, 2010
Digital DC-Reconstruction of AC-Coupled Electrophysiological Signals with a Single Inverting Filter
Roger Abächerli1,2,3, Jonas Isaksen1,4, Ramun Schmid1,3
1Signal Processing and Research, Research and Development, Schiller AG, Baar, Switzerland.
Abstract:
Since the introduction of digital electrocardiographs, high-pass filters have been necessary for successful analog-to-digital conversion with a reasonable amplitude resolution. On the other hand, such high-pass filters may distort the diagnostically significant ST-segment of the ECG, which can result in a misleading diagnosis. We present an inverting filter that successfully undoes the effects of a 0.05 Hz single pole high-pass filter. The inverting filter has been tested on more than 1600 clinical ECGs with one-minute durations and produces a negligible mean RMS-error of 3.1*10(-8) LSB. Alternative, less strong inverting filters have also been tested, as have different applications of the filters with respect to rounding of the signals after filtering. A design scheme for the alternative inverting filters has been suggested, based on the maximum strength of the filter. With the use of the suggested filters, it is possible to recover the original DC-coupled ECGs from AC-coupled ECGs, at least when a 0.05 Hz first order digital single pole high-pass filter is used for the AC-coupling.
More Related Videos
10:58A Procedure for Implanting Organized Arrays of Microwires for Single-unit Recordings in Awake, Behaving Animals
Published on: February 14, 2014
12:01Acute Dissociation of Lamprey Reticulospinal Axons to Enable Recording from the Release Face Membrane of Individual Functional Presynaptic Terminals
Published on: October 1, 2014
Related Concept Videos
Op Amp AC Circuits
Reconstruction of Signal using Interpolation
Half wave rectifier
Clamper Circuit
Within this circuit, the diode's orientation prompts the capacitor to charge up to the level of the most negative peak of the input signal. Upon reaching this state, the diode ceases to...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Clipper Circuit
The operation of a clipper circuit can be exemplified by analyzing a dual-clipper configuration setup that integrates two ideal diodes, each paired with a biasing...