Related Experiment Video
Updated: Mar 24, 2026

Ultrasonography of the Adult Male Urinary Tract for Urinary Functional Testing
Published on: August 14, 2019
Noninvasive Diagnosis of Bladder Outlet Obstruction in Patients with Lower Urinary Tract Symptoms Using Ultrasound
Muhammad Arif1, Jan Groen1, Egbert R Boevé2
1Department of Urology, Sector Furore, Erasmus Medical Centre, Rotterdam, The Netherlands.
Purpose:
We developed a noninvasive method to diagnose bladder outlet obstruction. An ultrasound based decorrelation method was applied in male patients with lower urinary tract symptoms.
Materials And Methods:
In 60 patients ultrasound data were acquired transperineally while they were voiding while sitting. Each patient also underwent a standard invasive pressure flow study.
Results:
High frequent sequential ultrasound images were successfully recorded during voiding in 45 patients. The decorrelation (decrease in correlation) between subsequent ultrasound images was higher in patients with bladder outlet obstruction than in unobstructed patients and healthy volunteers. ROC analysis resulted in an AUC of 0.96, 95% specificity and 88% sensitivity. A linear relationship was fitted to the decorrelation values as a function of the degree of obstruction represented by the bladder outlet obstruction index, measured in the separate pressure flow studies.
Conclusions:
It is possible to noninvasively diagnose bladder outlet obstruction using the ultrasound decorrelation technique.
Related Concept Videos
Imaging Studies II: Ultrasonography
Urodynamic Studies: Uroflowmetry
Imaging Studies VI: Voiding Cystourethrography and Cystography