Corneal biomechanical properties changes after coaxial 2.2-mm microincision and standard 3.0-mm phacoemulsification
Zhe Zhang1, Hua Yu2, Hui Dong2
1Clinical College of Ophthalmology, Tianjin Medical University, Tianjin Eye Hospital, Tianjin Key Laboratory of Ophthalmology and Visual Science, Tianjin Eye Institute, Tianjin 300070, China; Shanxi Eye Hospital, Taiyuan 030002, Shanxi Province, China.
Aim:
To compare the changes in corneal biomechanics measured by ocular response analyzer (ORA) after 2.2-mm microincision cataract surgery and 3.0-mm standard coaxial phacoemulsification.
Methods:
The prospective nonrandomized study comprised eyes with cataract that had 2.2-mm coaxial microincision or 3.0-mm standard incision phacoemulsification. The corneal hysteresis (CH), corneal resistance factor (CRF), corneal-compensated intraocular pressure (IOPcc) and Goldmann-correlated intraocular pressure (IOPg) were measured by ORA preoperatively and at 1d, 1-, 2-, 3- and 4-week postoperatively. Results were analyzed and compared between groups.
Results:
In both groups, CH decreased in the immediate postoperative period (P<0.05), returned to the preoperative level at one week (P=0.249) in the 2.2-mm group, and at two weeks in the 3.0-mm group (P=0.264); there was no significant change in CRF values. In 2.2-mm group, mean IOPcc and IOPg increased at 1d postoperatively (both P<0.05), and returned to preoperative level at one week (P=0.491 and P=0.923, respectively). In 3.0-mm group, mean IOPcc and IOPg increased at 1d and 1wk postoperatively (P=0.005 and P=0.029, respectively), and returned to preoperative level at 2wk (P=0.347 and P=0.887, respectively).
Conclusion:
Significant differences between preoperative and postoperative corneal biomechanical values were found for CH, IOPcc and IOPg. But the recovery time courses were different between the two groups. The 2.2-mm coaxial microincision cataract surgery group seemed recovery faster compared to the 3.0-mm standard coaxial phacoemulsification group.