Changes in intracuff pressure of cuffed endotracheal tubes while positioning for adenotonsillectomy in children
Griffin H Olsen1, Senthil G Krishna2,3, Kris R Jatana4
1The Ohio State University College of Medicine, Columbus, OH, USA.
Insights
Changes in head and neck positioning during pediatric adenotonsillectomy can alter cuffed endotracheal tube (cETT) intracuff pressure. Monitoring this pressure is crucial to prevent tracheal injury or airway complications.
Area of Science:
- Anesthesiology
- Pediatric Surgery
- Airway Management
Background:
- Cuffed endotracheal tubes (cETTs) are used in pediatric adenotonsillectomy.
- Head and neck positioning can significantly alter intracuff pressure in cETTs.
Purpose of the Study:
- To evaluate the combined impact of specific surgical positioning maneuvers on pediatric cETT intracuff pressure.
- Assessing risks associated with pressure fluctuations during adenotonsillectomy.
Main Methods:
- 84 pediatric patients undergoing adenotonsillectomy with cETTs were studied.
- Intracuff pressure was monitored continuously from baseline after cuff inflation.
- Pressure was re-evaluated after neck extension, shoulder roll placement, and Crowe-Davis retractor use.
Main Results:
- Intracuff pressure increased in 54.8% of patients post-positioning.
- 14.3% experienced pressures exceeding 30 cm H2O.
- Pressure decreased in 33.3% and remained unchanged in 11.9% of patients.
Conclusions:
- Pediatric adenotonsillectomy positioning can cause unpredictable intracuff pressure changes.
- High pressures risk tracheal mucosa damage; low pressures risk leaks and aspiration.
- Continuous or repeated intracuff pressure monitoring is recommended post-positioning.
Background:
When using cuffed endotracheal tubes (cETTs), changes in head and neck position can lead to changes in intracuff pressure.
Aim:
The aim of this study was to assess the combined effect of neck extension, shoulder roll placement, and Crowe-Davis retractor use during adenotonsillectomy on the intracuff pressure of cETTs in children.
Methods:
Patients <18 years of age undergoing adenotonsillectomy under general anesthesia following the placement of a cETT were included in the study. After inflation of the cuff to seal the trachea, using the leak test, baseline intracuff pressure was recorded and then continuously monitored. After neck extension, placement of a shoulder roll, insertion of the Crow-Davis retractor, suspension from a Mayo stand, and positioning for surgery, the intracuff pressure was recorded again.
Results:
The study cohort included 84 patients, ranging in age from 0.9 to 17 years (5.7 ± 3.9 years). In 46 patients (54.8%), the intracuff pressure increased from baseline after positioning for adenotonsillectomy. In 12 of these patients (14.3%), the intracuff pressure was >30 cm H2O. The intracuff pressure decreased in 28 patients (33.3%), while no change was noted in 10 patients (11.9%). Overall, the general trend was an increase in intracuff pressure from 15.9 ± 7.8 cm H2O to 18.9 ± 11.6 cm H2O.
Conclusion:
Both increases and decreases in the intracuff pressure may occur following positioning of the pediatric patient for adenotonsillectomy. An increase in intracuff pressure may result in a higher risk of damage to the tracheal mucosa. A decrease in the intracuff pressure can result in an air leak resulting in inadequate ventilation, increased risk of aspiration, and even predispose to airway fire if oxygen-enriched gases are used. Continuous intracuff pressure monitoring or rechecking the intracuff pressure after positioning for adenotonsillectomy may be indicated.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...