A noise-optimized virtual monochromatic reconstruction algorithm improves stent visualization and diagnostic accuracy
Stefanie Mangold1,2, Carlo N De Cecco1, U Joseph Schoepf3,4
1Department of Radiology and Radiological Science, Medical University of South Carolina, Ashley River Tower, 25 Courtenay Drive, Charleston, SC, 29425-2260, USA.
Purpose:
To evaluate the impact of noise-optimized virtual monochromatic imaging (VMI+) on stent visualization and accuracy for in-stent re-stenosis at lower extremity dual-energy CT angiography (DE-CTA).
Material And Methods:
We evaluated third-generation dual-source DE-CTA studies in 31 patients with prior stent placement. Images were reconstructed with linear blending (F_0.5) and VMI+ at 40-150 keV. In-stent luminal diameter was measured and contrast-to-noise ratio (CNR) calculated. Diagnostic confidence was determined using a five-point scale. In 21 patients with invasive catheter angiography, accuracy for significant re-stenosis (≥50 %) was assessed at F_0.5 and 80 keV-VMI+ chosen as the optimal energy level based on image-quality analysis.
Results:
At CTA, 45 stents were present. DSA was available for 28 stents whereas 12 stents showed significant re-stenosis. CNR was significantly higher with ≤80 keV-VMI+ (17.9 ± 6.4-33.7 ± 12.3) compared to F_0.5 (16.9 ± 4.8; all p < 0.0463); luminal stent diameters were increased at ≥70 keV (5.41 ± 1.8-5.92 ± 1.7 vs. 5.27 ± 1.8, all p < 0.001) and diagnostic confidence was highest at 70-80 keV-VMI+ (4.90 ± 0.48-4.88 ± 0.63 vs. 4.60 ± 0.66, p = 0.001, 0.0042). Sensitivity, negative predictive value and accuracy for re-stenosis were higher with 80 keV-VMI+ (100, 100, 96.4 %) than F_0.5 (90.9, 94.1, 89.3 %).
Conclusion:
80 keV-VMI+ improves image quality, diagnostic confidence and accuracy for stent evaluation at lower extremity DE-CTA.
Key Points:
• The impact of noise-optimized virtual monochromatic imaging on stent visualization was assessed. • Virtual monochromatic imaging significantly improves stent lumen visualization and diagnostic confidence. • At 80 keV diagnostic performance for detection of in-stent restenosis was increased. • 80 keV virtual monochromatic images are recommended for stent evaluation of lower extremity vasculature.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
04:35Sample Preparation for Computed Tomography-based Three-dimensional Visualization of Murine Hind-limb Vessels
Published on: October 7, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies VII: Vascular Imaging