Design of a 3T preamplifier which stability is insensitive to coil loading
Xueming Cao1, Elmar Fischer1, Jan G Korvink2
1Dept. of Radiology, Medical Physics, University Medical Center Freiburg, Freiburg, Germany.
None:
In MRI (magnetic resonance imaging), preamplifiers are needed to amplify signals obtained from MRI receiver coils. Under various loading conditions of the corresponding receiver coils, preamplifiers see different source impedance at their input and may become unstable. Therefore preamplifiers which stability is not sensitive to coil loading are desirable. In this article, a coil-loading-insensitive preamplifier for MRI is presented, derived from an unstable preamplifier. Different approaches to improve stability were used during this derivation. Since a very low noise factor is essential for MRI preamplifiers, noise contributions from passive components in the MRI preamplifier have to be considered during the stabilization process. As a result, the initially unstable preamplifier became stable with regard to coil loading, while other MRI requirements, as the extremely low noise factor, were still fulfilled. The newly designed preamplifier was manufactured, characterized and tested in the MRI spectrometer. Compared to a commercially available preamplifier, the newly designed preamplifier has similar imaging performance but other advantages like smaller size and better stability. Furthermore, presented stabilization approaches can be generalized to stabilize other unstable low-noise amplifiers.
More Related Videos
09:09Design and Construction of a Cost Effective Headstage for Simultaneous Neural Stimulation and Recording in the Water Maze
Published on: October 13, 2010
10:41A Wireless, Bidirectional Interface for In Vivo Recording and Stimulation of Neural Activity in Freely Behaving Rats
Published on: November 7, 2017
Related Concept Videos
Design Example: Vintage Mixing Console
The specifications for the pre-amplifier were clear. It needed to amplify the audio signal by a factor of 10, have an input impedance above 10...
Cascaded Op Amps
In a cascaded system, each op-amp is referred to as a stage. The output of one stage drives the input of the subsequent stage. As the input signal passes through...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Characteristics of Practical Op Amps
The ratio of differential gain to the common-mode gain is defined as the common-mode rejection ratio (CMRR). This ratio quantifies the ability of operational amplifiers (op-amps) to reject common-mode...
Inverting and Non-inverting OpAmps
Second-order Op Amp Circuits
The analysis of such circuits follows a systematic approach, similar to the second-order RLC circuits. In practical scenarios, bulky inductors are rarely employed due to their size and weight. This means...