Related Experiment Video
Updated: Mar 24, 2026

Evaluation of Fluid Overload by Bioelectrical Impedance Vectorial Analysis
Published on: August 17, 2022
Utility of stroke volume variation measured using non-invasive bioreactance as a predictor of fluid responsiveness in
Jeong Jin Min1, Jong-Hwan Lee2, Kwan Young Hong1
1Department of Anaesthesiology and Pain Medicine, Samsung Medical Centre, Sungkyunkwan University School of Medicine, 81 Irwon-ro, Gangnam-gu, Seoul, 06351, Korea.
Abstract:
The aim of this prospective study was to evaluate the usefulness of stroke volume variation (SVV) derived from NICOM® to predict fluid responsiveness in the prone position. Forty adult patients undergoing spinal surgery in the prone position were included in this study. We measured SVV from NICOM® (SVVNICOM) and FloTrac™/Vigileo™ systems (SVVVigileo), and pulse pressure variation (PPV) using automatic (PPVauto) and manual (PPVmanual) calculations at four time points including supine and prone positions, and before and after fluid loading of 6 ml kg-1 colloid solution. Fluid responsiveness was defined as an increase in the cardiac index from Vigileo™ of ≥12 %. There were 19 responders and 21 non-responders. Prone positioning induced a significant decrease in SVVNICOM, SVVVigileo, PPVauto, and PPVmanual. However, all of these parameters successfully predicted fluid responsiveness in the prone position with area under the receiver-operator characteristic curves for SVVNICOM, SVVVigileo, PPVauto, and PPVmanual of 0.78 [95 % confidence interval (CI) 0.62-0.90, P = 0.0001], 0.79 (95 % CI 0.63-0.90, P = 0.0001), 0.76 (95 % CI 0.6-0.88, P = 0.0006), and 0.84 (95 % CI 0.69-0.94, P < 0.0001), respectively. The optimal cut-off values were 12 % for SVVNICOM, SVVVigileo, and PPVauto, and 10 % for PPVmanual. SVV from NICOM® successfully predicts fluid responsiveness during surgery in the prone position. This totally non-invasive technique for assessing individual functional intravenous volume status would be useful in a wide range of surgeries performed in the prone position.
More Related Videos
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...