Optical Coherence Tomography: Imaging Mouse Retinal Ganglion Cells In Vivo
Published on: September 22, 2017
Comparison of Two Different OCT Systems: Retina Layer Segmentation and Impact on Structure-Function Analysis in
Livia M Brandao1, Anna A Ledolter2, Andreas Schötzau1
1Department of Ophthalmology, University of Basel, 4031 Basel, Switzerland.
Two spectral-domain optical coherence tomography (OCT) systems showed significant differences in macular thickness (MT) and ganglion cell layer-inner plexiform layer (GCIPL) measurements. However, both OCT systems effectively differentiated glaucoma patients from controls.
Area of Science:
- Ophthalmology
- Medical Imaging
- Glaucoma Research
Background:
- Spectral-domain optical coherence tomography (OCT) is crucial for glaucoma diagnosis.
- Comparing different OCT devices is essential for consistent clinical practice.
- Understanding structure-function relationships aids in glaucoma management.
Purpose of the Study:
- To compare macular thickness (MT) and ganglion cell-inner plexiform layer (GCIPL) measurements between Cirrus and Spectralis OCT systems.
- To evaluate the structure-function correlation between OCT parameters and standard automated perimetry (SAP) in glaucoma patients.
Main Methods:
- Seventeen primary open-angle glaucoma patients and 16 controls underwent OCT scans (Cirrus and Spectralis) and SAP testing.
- Differences in MT and GCIPL measurements between OCT devices were analyzed using t-tests and Bland-Altman analysis.
- Diagnostic performance was assessed via receiver operating characteristic (ROC) curves, and structure-function association was modeled using linear regression.
Main Results:
- Significant disagreement was found between Cirrus and Spectralis OCT for both MT and GCIPL measurements, with Spectralis yielding thicker values.
- Both OCT systems significantly differentiated glaucoma patients from controls for both MT and GCIPL.
- A significant negative association was observed between MT/GCIPL thickness and mean defect (MD) from SAP (p < 0.001).
Conclusions:
- While Cirrus and Spectralis OCT measurements are not interchangeable, both devices demonstrate significant diagnostic capability in distinguishing glaucoma patients.
- The structure-function correlation analysis yielded comparable results regardless of the OCT system used, supporting their utility in assessing glaucoma progression.