Related Experiment Video
Updated: Mar 24, 2026

A Magnetic Resonance Imaging Protocol for Stroke Onset Time Estimation in Permanent Cerebral Ischemia
Published on: September 16, 2017
Quiet diffusion-weighted head scanning: Initial clinical evaluation in ischemic stroke patients at 1.5T
Julie Rösch1, Martin Ott2,3, Bjoern Heismann2,4
1Department of Neuroradiology, University Hospital Erlangen, Erlangen, Germany. julie.roesch@uk-erlangen.de.
Quiet readout segmented echo-planar imaging diffusion-weighted imaging (q-DWI) offers comparable diagnostic value to routine ss-EPI-DWI for acute stroke MRI. q-DWI provides higher SNR and CNR, making it a suitable quiet alternative when acoustic noise is a concern.
Area of Science:
- Radiology
- Medical Imaging
- Neuroimaging
Background:
- Acute stroke diagnosis relies heavily on Magnetic Resonance Imaging (MRI).
- Diffusion-weighted imaging (DWI) is crucial for detecting acute ischemic lesions.
- Acoustic noise from MRI sequences can be a patient comfort issue.
Purpose of the Study:
- To compare the image quality and diagnostic performance of single-shot echo-planar imaging DWI (ss-EPI-DWI) with quiet readout segmented EPI-DWI (q-DWI).
- To evaluate q-DWI as a potentially quieter alternative for acute stroke MRI.
Main Methods:
- Twenty-six acute stroke patients underwent 1.5T MRI with both ss-EPI-DWI and q-DWI sequences.
- Quantitative (SNR, CNR) and qualitative (lesion conspicuity, artifacts, image quality, diagnostic content) assessments were performed.
- q-DWI was evaluated in both regular and fast acquisition modes.
Main Results:
- q-DWI demonstrated significantly higher signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) compared to ss-EPI-DWI.
- No significant difference was observed in lesion conspicuity between the sequences.
- ss-EPI-DWI was rated higher for overall image quality and had fewer artifacts than q-DWI.
- Diagnostic content was deemed equivalent between ss-EPI-DWI and q-DWI by radiologists.
Conclusions:
- q-DWI is a diagnostically equivalent alternative to ss-EPI-DWI for acute stroke MRI.
- q-DWI offers improved SNR and CNR, with significantly reduced acoustic noise.
- Consider q-DWI when patient comfort due to acoustic burden is a priority.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies VII: Vascular Imaging