A flexible pressure sensor could correctly measure the depth of chest compression on a mattress
Kouichiro Minami1, Yota Kokubo2, Ichinosuke Maeda2
1Emergency Life Saving and Technique Tokyo, Hachioji, Tokyo, Japan.
Background:
Feedback devices are used to improve the quality of chest compression (CC). However, reports have noted that accelerometers substantially overestimate depth when cardiopulmonary resuscitation (CPR) is performed on a soft surface. Here, we determined whether a flexible pressure sensor could correctly evaluate the depth CC performed on a mannequin placed on a mattress.
Methods:
Chest compression was performed 100 times/min by a compression machine on the floor or a mattress, and the depth of CC was monitored using a flexible pressure sensor (Shinnosukekun) and CPRmeter(™). The depth of machine-performed CC was consistently 5cm. We compared data from the feedback sensor with the true depth of CC using dual real-time auto feedback system that incorporated an infrared camera (CPR evolution(™)).
Results:
On the floor, the true depth of CC was 5.0±0.0cm (n=100), or identical to the depth of CC performed by the machine. The Shinnosukekun(™) measured a mean (±SD) CC depth of 5.0±0.1cm (n=100), and the CPRmeter(™) measured a depth of 5.0±0.2cm (n=100). On the mattress, the true depth of CC was 4.4±0.0cm (n=100). The Shinnosukekun(™) measured a mean CC depth of 4.4±0.0cm (n=100), and the CPRmeter(™) measured a depth of 4.7±0.1cm (n=100). The data of CPRmeter(™) were overestimated (P<.0001 between the true depth and the CPRmeter(™)-measured depth).
Conclusion:
The Shinnosukekun(™) could correctly measure the depth of CC on a mattress. According to our present results, the flexible pressure sensor could be a useful feedback system for CC performed on a soft surface.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:57Author Spotlight: Microfluidic Channel-Based Soft Electrodes and Their Application in Capacitive Pressure Sensing
Published on: March 17, 2023
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...