Consistency of corneal sublayer thickness measurements using Fourier-domain optical coherence tomography after
Alberto López-Miguel1, María Calabuig-Goena1,2, Victoria Marqués-Fernández2
1IOBA, Universidad de Valladolid, Valladolid - Spain.
Purpose:
To assess the reliability of corneal epithelial thickness (CET), nonepithelial central corneal thickness (NECCT), and central corneal thickness (CCT) measurements using Cirrus high-definition optical coherence tomography (HD-OCT) in patients who did and did not undergo cataract surgery.
Methods:
Forty patients who underwent uneventful phacoemulsification and 40 healthy participants were recruited to evaluate the intraobserver repeatability and interobserver reproducibility of CET, NECCT, and CCT measurements using Cirrus HD-OCT. To analyze repeatability, one examiner obtained 5 consecutive scans in each participant; for interobserver reproducibility, another examiner randomly obtained another scan. Within-subject standard deviation, coefficient of variation (CV), limits of agreement, and intraclass correlation coefficient (ICC) data were obtained.
Results:
For intraobserver repeatability, the intrasession CV (CVw) and ICC values of the CET in the operated and nonoperated groups were 3.7% and 0.80 and 3.8% and 0.73, respectively; for NECCT, 0.7% and 0.98 and 0.8% and 0.97; and for CCT, 0.6% and 0.99 and 0.7% and 0.98. For interobserver reproducibility, the CVw and ICC values for the CET in the operated and nonoperated groups were 2.6% and 0.82 and 2.3% and 0.62, respectively; for NECCT, 0.7% and 0.98 and 0.5% and 0.98; and for CCT, 0.5% and 0.99 and 0.4% and 0.99.
Conclusions:
The corneal sublayer thickness can be measured reliably using Cirrus HD-OCT in patients who underwent cataract surgery and elderly participants; however, the CET consistency is poorer than the NECCT. Corneal epithelial thickness modifications exceeding 4% reflect true thickness changes instead of random error variations using HD-OCT.