Performance of a non-contact infrared thermometer in healthy newborns
Sara Sollai1, Carlo Dani2, Elettra Berti1
1Department of Health Sciences, University of Florence, Anna Meyer Children's University Hospital, Florence, Italy.
Objective:
To evaluate the performance of a non-contact infrared thermometer (NCIT) in comparison with digital axillary thermometer (DAT) and infrared tympanic thermometers (ITT) in a population of healthy at term and preterm newborns nursed in incubators.
Setting:
1 level III maternity hospital, and its intensive neonatal care unit.
Participants:
119 healthy at term newborns and 70 preterm newborns nursed in incubators were consecutively enrolled. Exclusion criteria were unstable/critical conditions, polymalformative congenital syndromes and severe congenital syndromes.
Interventions:
Body temperature readings were prospectively collected. Each participant underwent bilateral axillary temperature measurement with DAT, bilateral tympanic measurement with ITT and mid-forehead temperature measurements using NCIT.
Primary Outcome Measures:
Degree of agreement between methods was evaluated by the Bland and Altman method.
Results:
714 measurements in 119 healthy at term newborns and 420 measurements in 70 preterm newborns nursed in incubators were performed. Clinical reproducibility of NCIT was 0.0455 °C for infants in incubators and 0.0861 °C for infants outside an incubator. Bias was 0.029 °C for infants in incubators and <0.0001 °C for infants outside an incubator. Zero outliers were recorded. The mean difference between methods was good both for newborns at term (0.12 °C for NCIT vs DAT and 0.02 °C for NCIT vs ITT) and preterm newborns in incubators (0.10 °C for NCIT vs DAT and 0.14 °C for NCIT vs ITT). Limits of agreement were 0.99 to -0.75 and 0.78 to -0.75 in at term newborns and were particularly satisfactory in preterm newborns in incubators (95% CI: 0.48 to -0.27 and 0.68 to -0.40).
Conclusions:
Our results with Bland and Altman analysis demonstrate that NCIT is a very promising tool, especially in preterm newborns nursed in incubators.
Trial Registration:
The study was approved by the Careggi University Hospital Ethics Committee (07/2011).
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...