Fetal heart rate monitoring of short term variation (STV): a methodological observational study
Stina Wretler1, Malin Holzmann2, Sophie Graner2,3
1Department of Women's and Children's Health, Karolinska Institutet, Karolinska University Hospital, 171 76, Stockholm, Sweden. stina.wretler@karolinska.se.
Background:
Cardiotocography (CTG) has high sensitivity, but less specificity in detection of fetal hypoxia. There is need for adjunctive methods easy to apply during labor. Low fetal heart rate short term variation (STV) is predictive for hypoxia during the antenatal period. The objectives of our study were to methodologically evaluate monitoring of STV during labor and to compare two different monitors (Sonicaid™ and EDAN™) for antenatal use.
Methods:
A prospective observational study at the obstetric department, Karolinska University hospital, Stockholm (between September 2011 and April 2015). In 100 women of ≥ 36 weeks gestation, STV values were calculated during active labor. In a subset of 20 women we compared STV values between internal and external signal acquisition. Additionally we compared antenatal monitoring with two different monitors in another 20 women.
Results:
Median STV in 100 fetuses monitored with scalp electrode during labor (EDAN™) was 7.1 msec (range 1.3-25.9) with no difference between early (3-6 cm) and late (7-10 cm) labor (7.1 vs 6.8 msec; p = 0.80). STV calculated from scalp electrode signals were positively correlated with delta-STV (STV internal -external) (R = 0.70; p < 0.01). No significant differences were found between Sonicaid™ and EDAN™ in antenatal external monitoring of STV (median difference 0.9 msec, Spearman Rank Correlation Sonicaid vs delta-STV; R = 0.35; p = 0.14).
Conclusions:
Median intrapartum STV was 7.1 msec. Significant differences were found between internal and external signal acquisition, a finding that suggests further intrapartum studies to be analysed separately depending upon type of signal acquisition. Antenatal external monitoring with Sonicaid™ and EDAN™ indicates that the devices perform equally well in the identification of acidemic fetuses. Further studies are needed to assess the clinical value of intrapartum STV.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessment of blood pressure in brachial artery(two-step method)