Influence of the X-ray beam quality on the dose increment in CT with iodinated contrast medium
Ernesto Amato1, Grazia Asero2, Salvatore Leotta3
1Department of Biomedical Sciences and of Morphologic and Functional Imaging, Section of Radiological Sciences, University of Messina, Messina, Italy.
Background:
In computed tomography (CT), the image contrast is given by the difference in X-ray attenuation in the various tissues of the patient and contrast media are used to enhance image contrast in anatomic regions characterized by similar attenuation coefficients.
Objective:
Aim of the present work is to enlarge the range of applicability of the method previously introduced for organ dosimetry in contrast-enhanced CT, by studying the effects of X-ray beam quality on the parameters of the model. Furthermore, an experimental method for the evaluation of the attenuation properties of iodinated solutions is proposed.
Methods:
Monte Carlo simulations of anthropomorphic phantoms were carried out to determine a bi-parametrical (a and b) analytical relationship between iodine concentration and dose increase in organs of interest as a function of the tube kilo-voltage peak potential (kVp) and filtration. Experimental measurements of increments in Hounsfield Units (HU) were conducted in several CT scanners, at all the kVp available, in order to determine the parameter γ which relates the HU increment with the iodine mass fraction. A cylindrical phantom that can be filled with iodine solutions provided with an axial housing for a pencil ionization chamber was designed and assembled in order to measure the attenuation properties of iodine solutions under irradiation of a CT scanner and to obtain a further validation of Monte Carlo simulations.
Results:
The simulation-derived parameters of the model, a and b, are only slightly dependent upon the tube kilo-voltage peak potential and filtration, while such scanner-dependent features influence mainly the experimentally-derived γ parameter. Relative dose variations registered by the ionization chamber inside the iodine-filled cylindrical phantom decrease when the X-ray mean energy increases, and reaches about 50% for 10 mg/ml of iodine.
Conclusions:
The dosimetric method for contrast-enhanced CT can be applied to all CT scanners by adopting average simulative parameters and by carrying out a simple measurement with a series of iodine contrast solutions. The novel experimental methodology introduced can provide a direct measurement of iodine attenuation properties.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT