High Frequency Ultrasound for the Analysis of Fetal and Placental Development In Vivo
Published on: November 8, 2018
Shear Wave Elastography in the Living, Perfused, Post-Delivery Placenta
Stephen A McAleavey1, Kevin J Parker2, Juvenal Ormachea2
1Department of Biomedical Engineering, University of Rochester, Rochester, New York, USA.
Abstract:
The placenta is the critical interface between the mother and the developing fetus and is essential for survival and growth. Despite the widespread use of ultrasound imaging and Doppler in obstetrics and gynecology and the recent growth of elastographic technologies, little is known about the biomechanical (elastic shear wave) properties of the placenta and the range of normal and pathologic parameters that are present. This study uses a well-developed protocol for perfusing whole placentas, post-delivery, to maintain tissue integrity and function for hours. In this model, the placenta is living, whole and maintained within normal physiologic parameters such as flow, arterial pressure and oxygen, throughout examination by ultrasound, Doppler and shear wave elastography. The preliminary results indicate that normal placental tissue on the fetal side has shear wave speeds on the order of 2 m/s, in a range similar to those of animal livers. Some abnormalities are found outside this range, and thus, elastographic measures of the placenta may provide useful assessments related to the state of the tissue.