Validation of simple and inexpensive algometry using sphygmomanometer cuff and neuromuscular junction monitor with
Padmaja Durga1, Sreedhar Reddy Wudaru1, Sunil Kumar Reddy Khambam2
1Department of Anesthesiology and Intensive Care, Nizam's Institute of Medical Sciences, Hyderabad, Andhra Pradesh, India.
Background And Aims:
The availability, ergonomics and economics prohibit the routine use of algometers in clinical practice and research by the anesthesiologists. A simple bedside technique of quantitative pain measurement would enable the routine use of algometry. We proposed to validate simple pain provocation using sphygmomanometer cuff and the electric stimulation of neuromuscular junction monitor (TOF-guard, Organon Teknika) to measure pain against a standardized laboratory pressure algometer.
Material And Methods:
Pain detection threshold (Pdt) and pain tolerance threshold (Ptt) were measured in forty healthy volunteers of both genders, using the above three techniques. All measurements were repeated three times. The co-efficient of inter-rater reliability (or consistency) between three independent measurements obtained from each of the techniques was determined by Cronbach's co-efficient alpha (α C). The correlation between the mean Pdt and Ptt values recorded by standardized algometer and the sphygmomanometer technique and nerve stimulator technique was performed using Pearson Correlation. An r >0.5 and a two-tailed significance of <0.05 were considered as good correlation between the standardized algometer and the tested techniques.
Results:
There was a good inter-rater reliability (α C > 0.7) for the three techniques. There was a good correlation with r >0.65 (P < 0.001) between the measurements of standardized pressure algometer and the two techniques being tested as alternatives for algometer to measure pain.
Conclusion:
The sphygmomanometer cuff technique and electrical stimulation with the peripheral nerve stimulator to measure pain threshold and tolerance provide a simple, efficient, repeatable measure of pain intensity and can be used as suitable alternatives to standard algometers.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure