Related Experiment Video
Updated: Mar 23, 2026

Author Spotlight: Assessing the Feasibility of Using Amplitude-Integrated EEG During Neonatal Transport
Published on: June 21, 2024
Physiological Signal Monitoring Bed for Infants Based on Load-Cell Sensors
Won Kyu Lee1, Heenam Yoon2, Chungmin Han3
1Interdisciplinary Program in Bioengineering, Graduate School, Seoul National University, Seoul 08826, Korea. wongyu86@snu.ac.kr.
Abstract:
Ballistocardiographs (BCGs), which record the mechanical activity of the heart, have been a subject of interest for several years because of their advantages in providing unobtrusive physiological measurements. BCGs could also be useful for monitoring the biological signals of infants without the need for physical confinement. In this study, we describe a physiological signal monitoring bed based on load cells and assess an algorithm to extract the heart rate and breathing rate from the measured load-cell signals. Four infants participated in a total of 13 experiments. As a reference signal, electrocardiogram and respiration signals were simultaneously measured using a commercial device. The proposed automatic algorithm then selected the optimal sensor from which to estimate the heartbeat and respiration information. The results from the load-cell sensor signals were compared with those of the reference signals, and the heartbeat and respiration information were found to have average performance errors of 2.55% and 2.66%, respectively. The experimental results verify the positive feasibility of BCG-based measurements in infants.