Comparison of tritiated-water standards by liquid scintillation for calibration of a new Standard Reference Material®
R Collé1, L Laureano-Perez1, D E Bergeron1
1National Institute of Standards and Technology, Gaithersburg, MD 20899-8462, USA.
Abstract:
A new National Institute of Standards and Technology (NIST) tritiated-water ((3)H-labeled oxidane) standard was prepared and calibrated. It is the 17th in a series of linked standards since 1954 and will be disseminated as Standard Reference Material® SRM 4927G, having a massic activity of 544.2kBqg(-1), with an expanded (k=2) relative standard uncertainty of 0.96%, at a Reference Time of 1200 EST, 1 May 2015. The calibration is based on relative liquid scintillation (LS) measurements using quench-varied efficiency tracing with two previous 1999 issues, viz., SRM 4927F and 4926E. Measurement comparisons were also made with respect to a 1994 tritiated-water French national standard and to a tritiated-water solution measured by 19 laboratories as part of an international measurement comparison organized by the Bureau International des Poids et Mesures (BIPM) in 2009. Confirmatory measurements for the massic activity of both SRM 4927F and 4927G by a triple-to-double coincidence ratio (TDCR) technique were also made.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
13:38Laser-Induced Fluorescence Emission L.I.F.E. as Novel Non-Invasive Tool for In-Situ Measurements of Biomarkers in Cryospheric Habitats
Published on: October 26, 2019
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Standard Solutions
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...