Comparison of disinfectants by immersion and spray atomization techniques on the linear dimensional stability of
Revathy Gounder1, B V J Vikas2
1Department of Prosthetic Dentistry, Faculty of Saveetha Dental College and Hospital, Saveetha University, Chennai, Tamil Nadu, India.
Objective:
To evaluate and compare the effect of 0.5% chlorhexidine gluconate, 1% sodium hypochlorite, and 2% glutaraldehyde by immersion and spray atomization technique on the linear dimensional stability of Jet bite, Aluwax and Ramitec interocclusal recording materials.
Materials And Methods:
Three representative materials: Jet bite (addition silicone), Aluwax and Ramitec (polyether) were mixed according to manufacturer's instructions and then specimens were prepared according to the specifications of ISO 4823. All the specimens except the control (distilled water) were treated with disinfectant solutions (0.5% chlorhexidine gluconate, 1% sodium hypochlorite, and 2% glutaraldehyde) for 30 and 60 min (n = 10) by spray and immersion technique. Once removed from the solutions, the test samples were washed in water for 15 s, dried and measured after 24 h 3 times using a measuring microscope with an accuracy of 0.0001 mm. Two-way ANOVA and Tukey's test with significance level of 5% were used to assess the statistical data (α = 0.05).
Result:
All groups showed no significant difference statistically, in linear dimension when disinfected for 30 min by spray or immersion technique. Polyether had significantly higher dimensional variation when immersed in sodium hypochlorite for 60 min. Addition silicone showed the least dimensional change which ranged from 0.024% to 0.05%, followed by polyether from 0.004% to 0.171% and Aluwax from 0.146% to 0.228%.
Conclusion:
To preserve the dimensions and surface of the recording materials and effective microbial elimination, restrictions should be applied in the method of disinfection and time duration. However, using the disinfectants either by spray or immersion technique, the dimensional change was <0.5% which was not clinically significant according to the American Dental Association specification no. 19 criteria within the first 24 h.
Related Concept Videos
In Vitro Drug Dissolution: Compendial Testing Models I
In Vitro Drug Dissolution: Alternative Methods
In Vitro Drug Dissolution: Compendial Testing Models II
Antimicrobial Effectiveness
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...