Related Experiment Video
Updated: Mar 23, 2026

Potentiodynamic Corrosion Testing
Published on: September 4, 2016
Medium-term results following large diameter metal-on-metal total hip arthroplasty: increasing failure after 6 years
Satish V Dhotare1, Deepak G Shivarathre1, Cristina Croitoru1
1Trauma and Orthopaedics Department, Royal Liverpool and Broadgreen University Hospitals NHS Trust, Liverpool - UK.
Purpose:
The main aim of our study is to report the medium-term survivorship of Birmingham Hip Resurfacing (BHR) cup and a large modular metal head (MMT) on an uncemented Freeman femoral stem. No results have been reported till date with these implants combinations.
Methods:
A total of 205 metal-on-metal total hip replacements (MoM THRs) were performed on 190 patients from October 2002 to November 2004. Prior to the Medicines and Healthcare Products Regulatory Agency (MHRA) guidelines, the patients were followed up at 3, 6 and 12 months postoperatively and annually thereafter. Following the MHRA alert in 2010, the patients were followed-up as per the MHRA guidelines.All statistical analyses were conducted using Stata 13.
Results:
At a mean follow-up of 10.5 years, a total of 42 out of 205 hips were revised for reasons related to ALTR. The failure rates increased significantly over time (7% at 6 years and 29% at 10-year follow-up). The analysis showed no statistically significant association to age, gender, side, BMI or component size or position (p<0.05). Blood metal ions showed a poor discriminant ability to separate failed from well-functioning MoM hip replacements.
Conclusions:
Large head MoM BHR on an uncemented stem has extremely high failure rate. The authors do not recommend the use of large head MoM articulation in total hip arthroplasty in the wake of this report and similar findings across the world. Continued surveillance of these implants is required as the failure rates increase with time.
More Related Videos
09:51The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
08:40Author Spotlight: A Novel 3D-Printed Titanium Implant for Minimally Invasive Treatment of Hip Dysplasia in Young Dogs
Published on: April 19, 2024