PSA discriminator influence on (222)Rn efficiency detection in waters by liquid scintillation counting
Ivana Stojković1, Nataša Todorović2, Jovana Nikolov2
1University of Novi Sad, Faculty of Technical Sciences, Novi Sad, Serbia.
Abstract:
A procedure for the (222)Rn determination in aqueous samples using liquid scintillation counting (LSC) was evaluated and optimized. Measurements were performed by ultra-low background spectrometer Quantulus 1220™ equipped with PSA (Pulse Shape Analysis) circuit which discriminates alpha/beta spectra. Since calibration procedure is carried out with (226)Ra standard, which has both alpha and beta progenies, it is clear that PSA discriminator has vital importance in order to provide precise spectra separation. Improvement of calibration procedure was done through investigation of PSA discriminator level and, consequentially, the activity of (226)Ra calibration standard influence on (222)Rn efficiency detection. Quench effects on generated spectra i.e. determination of radon efficiency detection were also investigated with quench calibration curve obtained. Radon determination in waters based on modified procedure according to the activity of (226)Ra standard used, dependent on PSA setup, was evaluated with prepared (226)Ra solution samples and drinking water samples with assessment of measurement uncertainty variation included.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Biological Effects of Radiation
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Emission Spectroscopy: Interference
Atomic Emission Spectroscopy: Lab