Repeatability in Color Measurements of a Spectrophotometer using Different Positioning Devices
Michael Hemming1, So Ran Kwon2, Fang Qian3
1Department of Dentistry, University of Iowa, College of Dentistry, Iowa City, USA.
Aim:
This study aimed to evaluate the repeatability of color measurements of an intraoral spectrophotometer with the use of three different methods by two operators.
Materials And Methods:
A total of 60 teeth were obtained, comprising 30 human maxillary teeth [central incisors (n = 10); canines (n = 10); molars (n = 10)] and 30 artificial teeth [lateral incisors (n = 10); premolar (n = 20)]. Multiple repeated color measurements were obtained from each tooth using three measuring methods by each of the two operators. Five typodonts with alternating artificial and human teeth were made. Measurements were taken by two operators with the Vita EasyShade spectrophotometer using the custom tray (CT), custom jig (CJ) and free hand (FH) method, twice, at an interval of 2 to 7 days. Friedman test was used to detect difference among the three color measuring methods. Post hoc Wilcoxon signed-rank test with Bonferroni correction applied was used for pair-wise comparison of color measurements among the three methods. Additionally, a paired-sample t-test was used to assess a significant difference between the two duplicated measurements made on the same tooth by the same operator for each color parameter and measuring method.
Results:
For operator A, mean (SD) overall color change-ΔE* (SD) perceived for FH, CT and CJ were 2.21(2.00), 2.39 (1.58) and 2.86 (1.92), respectively. There was statistically significant difference in perceived ΔE* in FH vs CJ (p = 0.0107). However, there were no significant differences between FH and CT (p = 0.2829) or between CT and CJ (p = 0.1159). For operator B mean ΔE* (SD) for FH, CT and CJ were 3.24 (3.46), 1.95 (1.19) and 2.45 (1.56), respectively. There was a significant difference between FH and CT (p = 0.0031). However, there were no statistically significant differences in ΔE* in FH vs CJ (p = 0.3696) or CT vs CJ (p = 0.0809).
Conclusion:
The repeatability of color measurements was different among the three measuring methods by operators. Overall, the CT method worked well for both operators.
Clinical Significance:
The use of a custom tray with apertures can improve the repeatability of color measurements of an intraoral spectrophotometer.
More Related Videos
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
UV–Vis Spectrometers
Uncertainty in Measurement: Accuracy and Precision
Uncertainty in Measurement: Reading Instruments