Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Assessment of cortical bone thickness using ultrasound
Katharina Degen1, Daniel Habor2, Klaus Radermacher2
1Department of Prosthodontics and Biomaterials, Medical Faculty, RWTH Aachen University, Aachen, Germany.
Purpose:
The aim of the study was to analyze the accuracy of measuring the cortical bone thickness using a combination of low- and high-frequency ultrasound (US) compared with cone-beam computed tomography (CBCT) and using stereomicroscopy as reference method.
Material And Methods:
Ten jawbone models were prepared using bovine ribs and porcine gingiva. A dental implant was placed in each model. All models were investigated by US, CBCT, and stereomicroscopy. The cortical bone thickness was measured directly above and 4 mm beside the implant with each method in different slices.
Results:
The median deviation of US measurements compared to the reference method was 0.23 mm. The CBCT method was slightly more accurate (median percent deviation of 9.2%) than the US method (10.3%). However, US measurements directly above the implant were more accurate than CBCT measurements with a median percent deviation of 10.5% for US vs. 11.8% for CBCT.
Conclusion:
Ultrasound showed a high potential to supplement CBCT for measurements of the cortical bone thickness.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):