Related Experiment Video
Updated: Mar 23, 2026

Conducting Respiratory Oscillometry in an Outpatient Setting
Published on: April 8, 2022
Validation of a metered dose inhaler electronic monitoring device: implications for asthma clinical trial use
Janine Pilcher1, Mark Holliday2, Stefan Ebmeier2
1Capital and Coast District Health Board, Wellington, New Zealand; Medical Research Institute of New Zealand, Wellington, New Zealand.
Background:
The SmartTouch Ventolin monitor (Adherium, Auckland, New Zealand) is an electronic monitor for use with a Ventolin metered dose inhaler, which records the date and time of inhaler actuations. This technology has the potential to allow in-depth analysis of patterns of inhaler use in clinical trial settings. The aim of this study was to determine the accuracy of the SmartTouch Ventolin monitor in recording Ventolin actuations.
Methods:
20 SmartTouch Ventolin monitors were attached to Ventolin metered dose inhalers. Bench testing was performed over a 10-week period, to reflect the potential time frame between visits in a clinical trial. Inhaler actuations were recorded in a paper diary, which was compared with data uploaded from the monitors.
Results:
2560 actuations were performed during the 10-week study period. Monitor sensitivity for diary-recorded actuations was 99.9% with a lower 97.5% confidence bound of 99.7%. The positive predictive value for diary-recorded actuations was 100% with a 97.5% lower confidence bound of 99.9%.
Conclusions:
The SmartTouch Ventolin monitor is highly accurate in recording and retaining electronic data. It can be recommended for use in clinical trial settings in which training and quality control systems are incorporated into study protocols to ensure accurate data acquisition.
Related Concept Videos
Inhaled Medications
Asthma-IV: Diagnostic and Management
Clinical Assessment for Asthma:
This is the first step in diagnosing and managing asthma. It includes:
Holter Monitor: 24-Hour Monitoring
Asthma-IV: Nursing Management
First, in...
Therapeutic Drug Monitoring: Drug Analysis Methods