Noninvasive pH monitoring for bacterial detection in platelet concentrates

Maria Loza-Correa1, Heather Perkins1, Dilini Kumaran1

  • 1Canadian Blood Services, Ottawa, Ontario, Canada.

Transfusion
|March 31, 2016
PubMed
Abstract

Insights

The pH SAFE system can detect bacterial contamination in platelet concentrates (PCs) by monitoring pH changes. A significant pH decrease indicates contamination, recommending continuous pH monitoring for quality control.

Area of Science:

  • Transfusion Medicine
  • Microbiology
  • Medical Device Technology

Background:

  • Bacterial contamination of platelet concentrates (PCs) is a major post-transfusion infectious risk.
  • The pH SAFE system offers a noninvasive method for PC quality control by measuring pH.

Purpose of the Study:

  • To evaluate the pH SAFE system as a rapid method for detecting bacterial contamination in PCs.
  • To assess the correlation between pH changes and bacterial load in stored PCs.

Main Methods:

  • Platelet concentrates were inoculated with various bacteria at low concentrations.
  • pH was measured noninvasively over 7 days using the pH SAFE system.
  • Bacterial concentration was determined through sample analysis.

Main Results:

  • A significant decrease in pH was observed in PCs with bacterial concentrations exceeding 10^7 CFUs/mL.
  • pH decline was detected as early as 28 hours for Bacillus cereus and 125 hours for Staphylococcus epidermidis.
  • Gram-negative bacteria showed an initial pH decline followed by a rebound.

Conclusions:

  • The pH SAFE system enables repeated, noninvasive pH screening of PCs during storage.
  • A significant pH decrease can indicate clinically relevant bacterial contamination.
  • Continuous pH monitoring is recommended due to species-specific pH decline patterns.