Positron detection in silica monoliths for miniaturised quality control of PET radiotracers
Mark D Tarn1, Dzmitry Maneuski, Richard Alexander
1Positron Emission Tomography Research Centre, University of Hull, Cottingham Road, Hull, HU6 7RX, UK. s.j.archibald@hull.ac.uk.
Abstract:
We demonstrate the use of the miniaturised Medipix positron sensor for detection of the clinical PET radiotracer, [(68)Ga]gallium-citrate, on a silica-based monolith, towards microfluidic quality control. The system achieved a far superior signal-to-noise ratio compared to conventional sodium iodide-based radio-HPLC detection and allowed real-time visualisation of positrons in the monolith.