Related Experiment Video
Updated: Mar 23, 2026

Robotic Ablation of Atrial Fibrillation
Published on: May 29, 2015
Clinical experience with contact-force and flexible-tip ablation catheter designs
Purpose:
Lesion formation is a critical determinant of technical and clinical success of pulmonary vein isolation. Different catheter designs aim to enhance tissue contact during ablation to enable optimized lesion formation. We analyzed procedural characteristics and predictors of clinical success in patients ablated with three different contemporary ablation catheters.
Methods:
Two hundred sixty-eight sequentially included patients receiving pulmonary vein isolation (PVI) with conventional (n = 122), contact-force (n = 96) and flexible-tip (n = 60) catheters were followed for a median of 14.1 months with 7d-Holter-monitoring and TTE at 3, 6, 12, and 24 months. Baseline characteristics and follow-up times were homogeneous across all groups.
Results:
Multivariable Cox proportional hazard regression for arrhythmia recurrence demonstrated a favorable hazard ratio for contact-force and flexible-tip catheters vs. conventional open irrigation catheters. Procedure time and fluoroscopy time were shorter for contact-force and flexible-tip catheters versus conventional catheters, but equal between. Linear lesions were applied in 58 % of contact-force and 66 % of flexible-tip cases, and CFAEs were targeted in 26 % of either.
Conclusions:
Our non-randomized prospectively collected data do not show a difference in observed procedure characteristics and in clinical outcome for flexible-tip versus contact-force catheter designs, while both display improved performance against conventional open irrigated-tip catheters. Linear lesions and CFAEs ablation were not associated with improved arrhythmia-free survival.