Electrogram Characteristics of Ablated and Non-Ablated Myocardium in Humans: A Comparison of Miniaturized Embedded
Michael S Lloyd1, Michael H Hoskins1, Anand D Shah1
1Emory University School of Medicine, Atlanta, Georgia, USA.
Introduction:
Intracardiac electrogram voltage remains an important metric for radiofrequency lesion application. Embedded micro-electrodes within ablation catheters are now approved for use in humans.
Objective:
This study examined electrogram characteristics of miniaturized electrode bipoles (iEGMmini ) compared to distal ablation tip bipoles (iEGMstandard ) in the setting of typical atrial flutter in humans.
Methods:
This was a single-center prospective trial involving 21 patients. A total of 54 non-ablated and 56 ablated sites were analyzed offline by separate blinded observers for iEGM voltage after a clinically successful radiofrequency (RF) ablation for typical atrial flutter. Ablation sites were defined as sites that were localized anatomically within the RF lesion set and did not have atrial capture at 10 mA and 2 milliseconds. Non-ablated sites were defined as sites adjacent to or remote from the ablation lesion set with capture at 10 mA and 2 milliseconds.
Results:
iEGMmini had greater voltages at non-ablated sites and lower voltages at ablated sites than iEGMstandard (non-ablated sites iEGMstandard 0.64 mV vs. iEGMmini 1.0 mV; ablated sites iEGMstandard 0.15 mV, iEGMmini 0.08 mV, P < 0.001). The difference for each patient between ablated and non-ablated sites was greater for iEGMmini compared to iEGMstandard (0.94 mV ± 0.57 iEGMmini vs. 0.52 mV ± 0.35 iEGMstandard , P < 0.001). Inter-observer agreement was good among all non-ablated sites and iEGMmini ablated sites, but was less robust for ablated sites using iEGMstandard .
Conclusions:
Imbedded mini-electrode iEGM bipoles had higher voltages in non-ablated tissue, lower voltages in ablated tissue, and significantly greater intra-patient differences between ablated and non-ablated sites. These data suggest a potential role for miniaturized electrode bipole use in RF lesion monitoring.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Electrophysiology of Normal Cardiac Rhythm
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...