Related Experiment Video
Updated: Mar 23, 2026

Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
Published on: August 4, 2023
Effects of cochlear implant surgical technique on post-operative electrode impedance
Ping Gu1, Yi Jiang1,2, Xue Gao1
1a Department of Otorhinolaryngology , Head and Neck Surgery, PLA General Hospital , Beijing , PR China ;
Conclusions:
The mCI surgical technique led to reduced impedance and minimized disturbance of the microenvironment inside the cochlea. Atraumatic surgical techniques and inflammation-reduction strategies may preserve the cochlear architecture and prevent fibrotic development.
Objectives:
To assess the advantages of a modified minimal access technique in cochlear implantation as well as to investigate the effects of intra-operative application of inflammation reduction strategies on the intra-cochlear impedance.
Methods:
One hundred patients received a 31.5 mm long Med-El standard electrode array fully inserted into the cochlea and did not have surgical complications post-operation. Patients were divided into two groups according to the surgical technique that was used for implantation: 50 were in the modified minimal access cochlear implant (mCI) surgery group and 50 were in the traditional CI surgery group. Intra-cochlear impedance values were measured at initial activation (4 weeks post-operatively). Electrode impedance values were compared between the surgery groups.
Results:
Electrode impedance values were significantly lower in the mCI group than in the CI surgery group at initial activation (5.01 kOhm vs 6.10 kOhm, respectively, F = 13.761, p = 0.000). The differences between the two groups were most prominent for the electrodes located at the basal region of the cochlea.