Raystation Monte Carlo application: evaluation of electron calculations with entry obliquity
Ben Archibald-Heeren1, Guilin Liu2
1Radiation Oncology Centre, Sydney, Australia. ben.archibald-heeren@roc.team.
Abstract:
To evaluate the accuracy of Raystation's implementation for Monte Carlo VMC ++ with electrons at varying angles of incidence for low and medium energy electron beams. Thirty-two profile and percentage depth dose scans were taken at 5° incident angle intervals for 6 and 12 MeV and compared to extracted fluences from Raystation calculations using gamma analysis with 2 %/2 mm criteria. Point dose measurements were compared to calculated doses to determine output accuracy. Electron profile and percentage depth dose curves for both energies show good agreement between 0° and 20° with 29/30 scans above 90 % pass rate. Average variation between calculated and measured point doses was -0.73 % with all measurements falling within ±2 % of calculated dose. Raystation's application of VMC ++ Monte Carlo algorithm provides clinically acceptable accuracy for low and medium energy electron dosimetry at incident angles up to 20° for Varian Clinac iX models.
Related Concept Videos
Electron Orbital Model
The first shell is closest to the nucleus, and it has only one subshell with a single spherical orbital called the...
π Electron Effects on Chemical Shift: Overview
Net Torque Calculations
Maxwell-Boltzmann Distribution: Problem Solving
This distribution function f(v) is defined by saying that the expected number N (v1,v2) of particles with speeds between v1 and v2 is given by
Calculation of Electric Flux
Magnetostatic Boundary Conditions