mARC prostate treatment planning with Varian Eclipse for flat vs. FFF beams
Katharina Bell1, Yvonne Dzierma1, Jan Palm1
1Department of Radiotherapy and Radiation Oncology, Saarland University Medical Centre, Kirrberger Str. Geb. 6.5, D-66421 Homburg/Saar, Germany.
Summary
The modulated arc (mARC) technique combined with flattening-filter-free (FFF) beams significantly reduces prostate cancer treatment time and out-of-field dose compared to traditional intensity-modulated radiation therapy (IMRT). This approach offers a preferable option for radiation oncology.
Area of Science:
- Radiation Oncology
- Medical Physics
Background:
- The modulated arc (mARC) technique is a rotational intensity-modulated radiation therapy (IMRT) delivery method utilizing burst mode.
- Varian's Eclipse treatment planning system (TPS) now includes an mARC planning option, necessitating systematic evaluation.
- Flattening-filter-free (FFF) beams offer potential advantages in radiation delivery efficiency and dose reduction.
Purpose of the Study:
- To systematically evaluate the mARC technique for prostate cancer treatment planning using the Eclipse TPS.
- To compare mARC with conventional IMRT using both 6MV and flattening-filter-free (FFF) 7MV beam energies.
- To assess plan quality, organ-at-risk sparing, out-of-field dose, and treatment delivery times.
Main Methods:
- Ten prostate cancer patients underwent standardized re-contouring and re-planning to 76Gy.
- Intensity-modulated radiation therapy (IMRT) and mARC plans were created using 6MV and FFF 7MV beams.
- Plans were compared based on plan quality indices, and delivered on an anthropomorphic phantom for dosimetry and time measurements.
Main Results:
- No significant difference in planning-target-volume (PTV) coverage between mARC and IMRT techniques.
- mARC demonstrated improved bladder sparing up to 75Gy; IMRT showed better sparing above this dose.
- The combination of FFF 7MV beams and mARC significantly reduced out-of-field dose and treatment time by a factor of three compared to 6MV IMRT.
Conclusions:
- All investigated modalities produced highly conformal treatment plans.
- The combination of high dose rate FFF beams with mARC is the preferable option for prostate cancer radiotherapy.
- This preferred approach offers significant reductions in treatment time and out-of-field dose.