In vitro biomechanical study of pedicle screw pull-out strength based on different screw path preparation techniques
Mark Moldavsky1, Kanaan Salloum1, Brandon Bucklen1
1Musculoskeletal Education and Research Center (MERC), A Division of Globus Medical, Inc., Audubon, PA 19403, USA.
Indian Journal of Orthopaedics
|April 8, 2016
Summary
Pedicle screw preparation techniques did not significantly alter screw-to-bone fixation strength. Despite variations in insertion torque, pull-out strength remained consistent across probed, under-tapped, standard-tapped, and sequential-tapped methods.
Area of Science:
- Spinal surgery
- Orthopedic biomechanics
- Spinal instrumentation
Background:
- Poor screw-to-bone fixation is a common clinical issue leading to screw loosening.
- Under-tapping (UT) pedicle screws has been previously studied biomechanically.
- The study aimed to assess if sequential tapping affects pedicle screw fixation strength.
Purpose of the Study:
- To evaluate the impact of different pedicle preparation techniques on screw-to-bone fixation strength.
- To compare under-tapping (UT), standard-tapping (ST), and sequential-tapping (SQT) methods.
- To utilize a stress relaxation protocol for assessing pull-out strength.
Main Methods:
- Three thoracolumbar calf spines were instrumented with pedicle screws.
- Pedicle preparation included probed, UT, ST, and SQT techniques.
- Pull-out strength was determined using a stress relaxation protocol; insertion torque was also measured.
Main Results:
- Insertion torques varied significantly between techniques (e.g., UT vs. ST/SQT).
- Pull-out strengths were 2443 N (probed), 2353 N (UT), 2474 N (ST), and 2146 N (SQT).
- No significant differences in pull-out strength were observed among the preparation techniques (P ≥ 0.05).
Conclusions:
- Standard-tapping (ST) yielded the highest insertion torque and pull-out strength, while sequential-tapping (SQT) showed the lowest.
- No significant differences in pedicle screw pull-out strength were found across tested preparation techniques.
- Bone density may be a more critical factor influencing screw fixation strength than preparation method.
Keywords:
Bone screwsCalf vertebral bodiesbiomechanicsfracture fixationinsertion torquepediclepull-out strengthspinal fracturesstress relaxationtapping technique