Purpose:
To investigate the relationship between the dynamic parameters (Ki) and static image-derived parameters of 68Ga-DOTATOC-PET, to determine which static parameter best reflects underlying somatostatin-receptor-expression (SSR) levels on neuroendocrine tumours (NETs).
Patients, Methods:
20 patients with metastasized NETs underwent a dynamic and static 68Ga-DOTATOC-PET before PRRT and at 7 and 40 weeks after the first administration of 90Y-DOTATOC (in total 4 cycles were planned); 175 lesions were defined and analyzed on the dynamic as well as static scans. Quantitative analysis was performed using the software PMOD. One to five target lesions per patient were chosen and delineated manually on the baseline dynamic scan and further, on the corresponding static 68Ga-DOTATOC-PET and the dynamic and static 68Ga-DOTATOC-PET at the other time-points; SUVmax and SUVmean of the lesions was assessed on the other six scans. The input function was retrieved from the abdominal aorta on the images. Further on, Ki was calculated using the Patlak-Plot. At last, 5 reference regions for normalization of SUVtumour were delineated on the static scans resulting in 5 ratios (SUVratio).
Results:
SUVmax and SUVmean of the tumoural lesions on the dynamic 68Ga-DOTATOC-PET had a very strong correlation with the corresponding parameters in the static scan (R²: 0.94 and 0.95 respectively). SUVmax, SUVmean and Ki of the lesions showed a good linear correlation; the SUVratios correlated poorly with Ki. A significantly better correlation was noticed between Ki and SUVtumour(max and mean) (p < 0.0001).
Conclusions:
As the dynamic parameter Ki correlates best with the absolute SUVtumour, SUVtumour best reflects underlying SSR-levels in NETs.
More Related Videos
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...