Related Experiment Video
Updated: Mar 22, 2026

A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
Pressure attenuation during high-frequency airway clearance therapy across different size endotracheal tubes: An in
Craig D Smallwood1, Kevin J Bullock2, Andrew Gouldstone3
1Division of Critical Care Medicine, Department of Anesthesiology, Perioperative and Pain Medicine, Boston Children's Hospital, Boston, MA; Department of Bioengineering, Northeastern University, Boston, MA; Harvard Medical School, Boston, MA.
Purpose:
High-frequency airway clearance therapy is a positive pressure secretion clearance modality used in pediatric and adult applications. However, pressure attenuation across different size endotracheal tubes (ETT) has not been adequately described. This study quantifies attenuation in an in vitro model.
Materials And Methods:
The MetaNeb® System was used to deliver high-frequency pressure pulses to 3.0, 4.0, 6.0 and 8.0mm ID ETTs connected to a test lung during mechanical ventilation. The experimental setup included a 3D-printed trachea model and imbedded pressure sensors. The pressure attenuation (Patt%) was calculated: Patt%=[(Pproximal-Pdistal)/Pproximal]x100. The effect of pulse frequency on Pdistal and Pproximal was quantified.
Results:
Patt% was inversely and linearly related to ETT ID and (y=-7.924x+74.36; R(2)=0.9917, P=.0042 for 4.0Hz pulse frequency and y=-7.382+9.445, R(2)=0.9964, P=.0018 for 3.0Hz pulse frequency). Patt% across the 3.0, 4.0, 6.0 and 8.0mm I.D. ETTs was 48.88±10.25%, 40.87±5.22%, 27.97±5.29%, and 9.90±1.9% respectively. Selecting the 4.0Hz frequency mode demonstrated higher Pproximal and Pdistal compared to the 3.0Hz frequency mode (P=.0049 and P=.0065). Observed Pdistal was <30cmH2O for all experiments.
Conclusions:
In an in vitro model, pressure attenuation was linearly related to the inner diameter of the endotracheal tube; with decreasing attenuation as the ETT size increased.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...