Interlocking Nailing Versus Interlocking Plating in Intra-articular Calcaneal Fractures: A Biomechanical Study
Sophia Reinhardt1, Heiner Martin2, Benjamin Ulmar3
1Department of Trauma, Rostock University Medical Center, Hand and Reconstructive Surgery, Rostock, Germany sophia.labs@gmx.de.
Minimally invasive calcaneal interlocking nails offer comparable primary stability to traditional plates for displaced intra-articular calcaneal fractures. The C-Nail demonstrated superior performance in dynamic failure testing compared to the Calcanail and Rimbus plate.
Area of Science:
- Orthopedic surgery
- Biomechanics
- Trauma surgery
Background:
- Open reduction and internal fixation with plates is the standard for calcaneal fractures, but carries high risks of wound complications.
- Minimally invasive techniques offer potential benefits but may compromise construct stability.
- Two novel interlocking nail systems were developed for minimally invasive treatment with enhanced stability.
Purpose of the Study:
- To compare the primary stability of two minimally invasive calcaneal interlocking nail systems against a variable-angle interlocking plate.
- To evaluate biomechanical performance under dynamic and load-to-failure conditions.
Main Methods:
- A standardized Sanders type IIB calcaneal fracture model was created in 21 cadavers.
- Osteosynthesis was performed using two interlocking nail systems (C-Nail, Calcanail) and one interlocking plate (Rimbus).
- Biomechanical testing included dynamic load sequences and load-to-failure tests, with interfragmentary movement and Boehler angle measurements.
Main Results:
- No significant differences in load to failure, stiffness, or Boehler angle were observed between fixation systems.
- The C-Nail exhibited the highest load to failure.
- Dynamic failure testing revealed significantly higher failure rates for Calcanail (87.5%) and Rimbus plate (66%) compared to C-Nail (14%).
Conclusions:
- Both minimally invasive interlocking nail systems demonstrate high primary stability, comparable to interlocking plates.
- Interlocking nails are a viable alternative for treating displaced intra-articular calcaneal fractures, with C-Nail showing promising results in dynamic stability.